Abstract
In this work we have characterised the K-edge digital subtraction angiography imaging system at the European Synchrotron Radiation Facility (ESRF, Grenoble, France). With this technique, after the minimally invasive intravenous injection of a contrast agent (iodine), two images are recorded simultaneously using monochromatic beams with energies bracketing the iodine K-edge. The image receptor consists of a two-line 432-pixel germanium detector. The logarithmic subtraction of the image set produced results in an iodine-enhanced image, where vessels are clearly visualised and contrast agent concentration can be precisely quantified. We have studied the imaging system in terms of the modulation transfer function, which was measured at the patient position, the 2-D normalised noise power spectrum, calculated for both raw and processed data, and the frequency-dependent detective quantum efficiency, which was calculated directly for final images. In addition, images of cylindrical phantoms with different concentrations of iodine, were also acquired.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.