Abstract
High complication rates and surgical revision rates following Aptis implant placement have been reported in the literature. This study evaluates the performance of the Aptis implant of twelve patients using four-dimensional kinematic analysis. The (mean) follow-up was 58 months. Wrist motion, grip strength, and kinematic analysis of both arms were used to investigate possible causes of the reported complications. In nine cases (75%), the proximal to distal translation of the distal radius along the ulnar axis in the affected forearm was too little or absent. Significant correlations were found between postoperative extension and translation of the distal radius along the ulnar axis and between the radial deviation and combined error. The four-dimensional kinematic analysis suggests that the current design of the implant could lead to limited restoration of the position of the forearm rotation axis and the translation of the radius along the ulnar axis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.