Abstract

• Calibration of the Xpert with viral stock dilutions was linear and reproducible. • Overall Xpert/ID-NOW positive agreement = 70% and negative agreement = 99.6%. • Sensitivity of ID-NOW for Xpert N2 CT ≤ 37 (viral loads ≥ 945 copies/mL) was 100%. • Sensitivity of ID-NOW for Xpert N2 for N2 CT > 37 was only 37.7%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.