Abstract

A Small- and Wide-Angle X-ray Scattering (SWAXS) beamline (BL-18) is installed and commissioned at a 1.5 T bending magnet port (5°) of Indus-2 synchrotron at RRCAT, Indore, India. The ∼40-m-long beamline has tunable x-ray energy in the range of 5–20 keV by using a double crystal monochromator. A 1.5-m-long toroidal mirror is used to focus the x-ray beam at the detector position. The beamline is equipped with a 6-m-long movable detector stage to access different wave-vector transfer ranges. At present, an online image plate area detector and a linear position-sensitive gas detector are installed for Small-Angle X-ray Scattering (SAXS) and Wide-Angle X-ray Scattering (WAXS) measurements, respectively. The beamline is operational in simultaneous SAXS/WAXS mode to probe the mesoscopic as well as molecular level structure over a wide range of wave-vector transfer. The specification of the beamline and its performance are reported here. A few recent experimental results, as obtained from BL-18, are also described in brief.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call