Abstract

Active personal dosemeters (APDs) are very useful instruments for optimizing radiation protection of workers and for increasing worker’s awareness of unexpected exposures. The challenge of monitoring personal equivalent doses with APDs in interventional fluoroscopy is that they must be sensitive to low energy photon beams and be able to record high dose rates. The aim of this work is to verify both the performance and the reliability of four active personal dosemeters (APDs) and one direct ion storage (DIS) dosemeter in typical X-ray radiation fields used during interventional radiology and cardiology (IR/IC) procedures. The values of the personal dose equivalent at a depth of 10 mm measured by the APDs are compared with the response of a whole body thermoluminescence dosemeter (TLD) as a reference dosemeter. The response is found to be satisfactory in the tested situations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.