Abstract
The influence of up to 16% HNO3, 28% HCl, and the mixture of both acids in aqua regia on the analytical performance of electrodeposited modifiers (Ir or Ir+Pd) was evaluated and discussed. Cadmium was used as an example of volatile elements often determined by graphite furnace atomic absorption spectrometry (GF AAS). In the presence of HCl, the maximum pyrolysis temperature that could be applied was found to be 600 degrees C. In the presence of HNO3 and aqua regia, both modifiers stabilized cadmium up to 800 degrees C. The long-term performance of electrodeposited Ir or Ir+Pd was not influenced by mineral acids; moreover the tube lifetime was significantly prolonged compared with a non-modified tube.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.