Abstract

The paper evaluates the performances of nonylphenol ethoxylates free cocktail (ProSafe LT+, Meridian Biotechnologies Ltd.) for tritium measurement using liquid scintillation counting and the laboratory’s routine procedure. The optimum ratio of sample and liquid scintillation cocktail for 20 ml vial volume was investigated, choosing a ratio of 8:12 ml of sample: scintillation cocktail. Good results were obtained for linearity (investigated between 0.3 MBq l−1 and 6 Bq l−1), repeatability, and trueness, and prove its applicability both for operation and environmental monitoring of tritium at nuclear facilities.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.