Abstract

Background With the development of liquid biopsy technology, the demand for noninvasive prenatal testing (NIPT) is increasing rapidly. The aim of the study is to evaluate the effects of different blood collection tubes on plasma cfDNA and NIPT quality control. Methods We investigated hemolysis, cfDNA concentration, and fragment distribution within blood samples stored in EDTA, ImproGene, and Streck tubes. The effects of ImproGene and Streck tubes on NIPT quality control were evaluated. Results The ImproGene tubes prevented the time-dependent increase of cfDNA concentration and preserved the cfDNA fragment size distribution. For NIPT quality control, there is no significant difference in cfDNA, library concentration, and fetal fraction between ImproGene and Streck tubes samples. GC content of the samples in ImproGene tubes was closer to the human genome. Conclusion The ImproGene cfDNA tube has excellent performance and is an effective choice for storing blood samples for NIPT testing or other cfDNA analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.