Abstract
The performance of image intensifier (II)-equipped C-arm systems for three-dimensional (3D) imaging was investigated. The three-dimensional image quality was evaluated in terms of spatial resolution (modulation transfer function, MTF), contrast resolution, geometrical accuracy and homogeneity depending on the image intensifier format, focal spot size, number of projections and the angular span covered during the data acquisition. Experiments have been performed on a vascular C-arm system equipped with a 38-cm image intensifier. Several objects, including CT performance, MTF and pelvis phantoms, were scanned under various conditions. It was shown that for reasonable acquisition parameters, a contrast resolution below 100 HU could be obtained with standard acquisition strategies. Focusing on the spatial resolution, an almost isotropic three-dimensional resolution of up to 22 lp/cm at 10% modulation could be obtained when a 17-cm II was used. The homogeneity in the resulting images was limited by the remaining scatter and truncations. The resulting geometrical accuracy was in the order of the voxel size.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.