Abstract

This paper investigates, through numerical simulations, the performance of a signal analysis method by which a high temperature fission chamber can be used over a wide range of count rates. Results reported in a previous paper (Elter et al., 2015 [1]) indicated that the traditional Campbell method and the pulse mode cannot provide a sufficient overlap at medium count rates. Hence the use of the so-called Higher Order Campbell (HOC) methods is proposed and their performance is investigated.It is shown that the HOC methods can guarantee the linearity (i.e. correctness) of the neutron flux estimation over a wide count rate, even during transient conditions. The capabilities of these methods for suppressing parasitic noise (originating from various sources) are verified.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.