Abstract

Rapid antigen tests are commonly used by clinicians for rapid, simple, point-of-care testing. Five rapid antigen tests were shown to have low sensitivity (40.3-58.8%) when compared to real-time RT-PCR using nasal wash specimens from patients with influenza-like-illness (N = 167) that were collected previously and confirmed as 2009 pandemic influenza A (H1N1)-positive by PCR. Rapid antigen test sensitivity correlated with virus levels in nasal secretions when comparisons were made to cycle threshold (C(T)) values obtained from real-time RT-PCR. When C(T) values are <25 (equating to viral concentrations of >10(4) TCID(50)/ml) sensitivity for all five rapid antigen kits was high (range: 83-94% positive); however, when C(T) values are >30 (10(2) TCID(50)/ml), sensitivities of only 16-18% were observed for four of five rapid antigen kits. The Directigen EZ Flu A + B test detected more positive samples (35%) at lower viral concentrations with C(T) values >30 when compared with other commercial kits (P = 0.05). Rapid antigen test results must be interpreted with caution, and negative specimens may need confirmation by sensitive molecular assays.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.