Abstract

The performance of the extrapolated leading edge type of time pickoff used with a large HPGe coaxial detector is studied. The time resolution using such a type of time pickoff has been found to be comparable with that obtained by employing the commonly used amplitude and rise-time compensation technique. For a BaF 2−HPGe (27% efficiency) coincidence setup, the values of FWHM and FWTM are 3.2 and 7.0 ns, respectively, when 60Co photopeaks are selected.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.