Abstract

BackgroundThis study was designed to evaluate emergency physician success and satisfaction using a video-assisted semi-rigid fiberoptic stylet, the Clarus Video System (CVS), during a simulated difficult airway scenario.FindingsEmergency physicians (EPs) of all levels were first shown a brief slide show and three example videos, and then given 20 min to practice intubating a mannequin using both the CVS and standard direct laryngoscopy (DL). The mannequin was then placed in a c-collar and set to simulate an apneic patient with an edematous tongue and trismus. Each EP was given up to three timed attempts with each technique. They rated their satisfaction with the CVS, usefulness for their practice, and the effectiveness of the tutorial. Direct laryngoscopy had a 65% success rate on the first attempt, 20% on the second, and 15% required three or more. The CVS had a 100% success rate with a single attempt. Average time for independent DL attempts was 43.41 s (SD = ±26.82) and 38.71 s (SD = ±34.14) with CVS. Cumulative attempt times were analyzed and compared (DL = 74.55 ± 68.40 s and CVS = 38.71 ± 34.14 s; p = 0.028). EPs rated their satisfaction with, and usefulness of, the CVS as ≥6 out of 10.ConclusionEmergency physicians were able to successfully intubate a simulated difficult airway model on the first attempt 100% of the time. Emergency physicians were satisfied with the CVS and felt that it would be useful in their practice.

Highlights

  • This study was designed to evaluate emergency physician success and satisfaction using a video-assisted semi-rigid fiberoptic stylet, the Clarus Video System (CVS), during a simulated difficult airway scenario

  • During the practice period the mannequin was set to simulate an apneic patient with normal airway, and no restrictions were imposed upon technique or positioning of the head or neck

  • One emergency physician (EP) failed to intubate with direct laryngoscopy (DL) after three attempts

Read more

Summary

Conclusion

Emergency physicians were able to successfully intubate a simulated difficult airway model on the first attempt 100% of the time.

Background
Methods
Results
Discussion
Conclusions
Levitan RM
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.