Abstract

The EasyCell assistant (Medica, Bedford, MA, USA) is one of the state-of-the-art digital morphology analyzers. We explored the performance of EasyCell assistant in comparison with manual microscopic review and Pentra DX Nexus (Horiba ABX Diagnostics, Montpellier, France). In a total of 225 samples (100 normal and 125 abnormal samples), white blood cell (WBC) differentials and platelet (PLT) count estimation by EasyCell assistant were compared with the results by manual microscopic review and Pentra DX Nexus. The manual microscopic review was performed according to the Clinical and Laboratory Standards Institute guidelines (H20-A2). WBC differentials between pre-classification by EasyCell assistant and manual counting showed moderate correlations for neutrophils (r=0.58), lymphocytes (r=0.69), and eosinophils (r=0.51) in all samples. After user verification, they showed mostly high to very high correlations for neutrophils (r=0.74), lymphocytes (r=0.78), eosinophils (r=0.88), and other cells (r=0.91). PLT count by EasyCell assistant highly correlated with that by Pentra DX Nexus (r=0.82). The performance of EasyCell assistant for WBC differentials and PLT count seems to be acceptable even in abnormal samples with improvement after user verification. The EasyCell assistant, with its reliable performance on WBC differentials and PLT count, would help optimize the workflow of hematology laboratories with reduced workload of manual microscopic review.