Abstract

A digital gamma-ray spectrometer with loss-free counting system was used for real time correction for dead time (DT) due to pulse processing in the electronics. This spectrometer was used for radioactive assay of samples having constant as well as varying DT conditions. The system was tested by measuring activity of short and medium lived nuclides namely 28Al, 52V and 128I in the DT ranges of 80–2 %. Using this spectrometer and neutron activation analysis (NAA), concentrations of Al, V, Ti, Ca, Dy and Mn were determined in some samples and reference materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.