Abstract
Performance of cerium-doped Gd3Al2Ga3O12 (GAGG:Ce) scintillator in gamma-ray spectrometry has been investigated. The measurements of two samples of GAGG:Ce cover the tests of emission spectra (maximum of emission at about 530nm), light output, non-proportionality, energy resolution, time resolution and decay time of light pulses. We compare the results with commonly known scintillators, such as NaI(Tl), LSO, LuAG etc. The results show that GAGG:Ce has a high light yield of about 33000ph/MeV as measured with Hamamatsu S3590–18 Si PiN photodiode [1]. The total energy resolution for 662keV gamma-rays from 137Cs source is equal to about 6%, whereas intrinsic resolution is equal to 5.2%. Additionally, we made basic measurements of photoelectron yield, non-proportionality and total energy resolution of small sample (5×5×5mm3) of GAGG:Ce crystal coupled to Hamamatsu MPPC array (6×6mm2). The results show that the performance of GAGG:Ce measured with the MPPC array are similar to those measured with the PMT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.