Abstract
Bhabha Atomic Research Centre (BARC) conducts fast neutron personnel monitoring service to radiation workers involved in reactors, accelerators, spent fuel processing plants, oil-well industries, etc., using CR-39 detector based dosemeter. In this study, performance of the BARC fast neutron personnel dosemeter has been checked through EURADOS intercomparison exercise for simulated workplace neutron fields. The overall performance of the dosemeter in the lower dose equivalent (≤ 5mSv) was found to be acceptable as per ISO-14146. The performance (ratio of estimated to reference dose equivalent) of the dosemeter in the higher dose equivalent (12mSv) was found to be in the range of 0.48-0.44 and not satisfactory as per the ISO-14146 criteria for any dosimetry service. Based on this performance, a new imaging system was developed and performance of the dosemeters were improved and found to be acceptable (within ±20%) as per ISO.
Accepted Version
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.