Abstract

Retrospective analysis of volumetric modulated arc therapy plans for hypofractionated left whole-breast irradiation with simultaneous integrated boost to assess the performance of the auto-planning (AP) engine. Fifteen treatment plans, produced using manual optimization planning approach (MP) were replanned using (AP) approach. Dose-volume parameters were defined to quantify the quality of concurrent treatment plans assessing target coverage and sparing organs at risk (OARs). The Wilcoxon Signed Rank test was used for statistical comparison of all results obtained from the use of the 2 approaches. Dose coverage for both PTVs, PTVbreast, and PTVboost, were similar with AP showing slightly significantly better results for the homogeneity index for both PTVs, for D98% of PTVbreast and D2% of PTVboost. AP plans provided a significant reduction of dose for ipsilateral lung and contralateral lung. No significant differences were observed for heart and contralateral breast. A percentage difference of -14.0% was found for the mean dose of left coronary artery between AP plans and MP plans. Despite increase of total MU by 4.3% for AP plans, planning time resulted about half of that of the MP approach. Although PTVs doses were similar between MP and AP plans, AP plans generally spared OARs significantly better than MP plans. Furthermore, the shortest AP treatment plan time approach was attractive with respect to the workload.