Abstract

Simple methods of quantifying selected physical characteristics of angiographic catheters, which are likely to influence their in vivo performance, have been developed. The torsional stiffness ('torque') of catheters was measured, and tip control ('manoeuvrability') in an angiographic phantom was studied. Both braided and unbraided catheters were investigated all having the 'headhunter' tip shape. Significant differences between measurements made at 20 and 37 degrees C were found. Torsional stiffness was not a good predictor of performance in the angiographic phantom. Ethylene oxide re-sterilization has only a limited effect on the physical characteristics studied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.