Abstract
To assess the performance and to provide more detailed insight into the characteristics and limitations of devices for automatic dose control (ADC) in CT. A comprehensive study on DoseRight 2.0, the ADC system provided by Philips for its Brilliance CT scanners, was conducted. Phantom tests were carried out on a 64-slice scanner (Brilliance 64) using assorted quality control (QC) phantoms that allowed verification of the basic specifications. If feasible, the findings were verified by model calculations based on known specifications. For all tests, the dose reductions and modulation characteristics fully met the values expected from the specifications. Adverse effects due to increased image noise were only moderate as a result of the "adequate noise system" design that employs comparatively gentle modulation, and the additional use of adaptive filtration. Simple tests with QC phantoms allow evaluation of the most relevant characteristics of devices for ADC in CT.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.