Abstract

Abstract IgG antibodies coating red blood cells (RBCs) can be removed by elution procedures and their specificity determined by antibody identification studies. Although such testing is traditionally performed using the tube agglutination assay, prior studies have shown that the gel microcolumn (GMC) assay may also be used with comparable results. The purpose of this study was to compare an automated solid-phase red cell adherence (SPRCA) system with a GMC assay for the detection of antibodies eluted from RBCs. Acid eluates from 51 peripheral blood (PB) and 7 cord blood (CB) samples were evaluated by both an automated SPRCA instrument and a manual GMC assay. The concordance rate between the two systems for peripheral RBC samples was 88.2 percent (45 of 51), including cases with alloantibodies (n = 8), warm autoantibodies (n = 12), antibodies with no identifiable specificity (n = 2), and negative results (n = 23). There were six discordant cases, of which four had alloantibodies (including anti-Jka, -E, and -e) demonstrable by the SPRCA system only. In the remaining 2 cases, anti-Fya and antibodies with no identifiable specificity were demonstrable by the GMC assay only. All seven CB specimens produced concordant results, showing anti-A (n = 3), -B (n = 1), maternal anti-Jk a (n = 2), or a negative result (n = 1). Automated SPRCA technology has a performance that is comparable with that of a manual GMC assay for identifying antibodies eluted from PB and CB RBCs. Immunohematology 2011;27:1–5.