Abstract

High-resolution gamma-ray detectors based on high-purity germanium crystals (HPGe) are one of the key workhorses of experimental nuclear science. The technical development of such detector technology has been dramatic in recent years. Large volume, high-granularity, electrically segmented HPGe detectors have been realised and a methodology to improve position sensitivity using pulse-shape analysis coupled with the novel technique of gamma-ray tracking has been developed. Collaborations have been established in Europe (Advanced GAmma Tracking Array (AGATA)) [J. Simpson, Acta Phys. Pol. B 36 (2005) 1383] and the USA (GRETA/GRETINA) [C.W. Beausang, Nucl. Instr. and Meth. B 204 (2003)] to build gamma-ray tracking spectrometers. This paper discusses the performance of the first AGATA asymmetric detector that has been tested at the University of Liverpool. The use of a fully digital data acquisition system has allowed detector charge pulse shapes from a selection of well-defined photon interaction positions to be analysed, yielding important information on the position sensitivity of the detector.