Abstract
ABSTRACT The performance of a scanning differential mobility analyzer (DMA) combined with a condensation nucleus counter (CNC) having high detection sensitivity has been evaluated using monodisperse latex particles of 100 nm in diameter and polydisperse sodium chloride particles between 10–120 nm: 1) A method is proposed to accurately measure the delay time between the DMA outlet and the particle detector of the CNC, which is necessary to measure the particle size distribution by the scanning DMA. 2) The effect of some operating factors, such as CNC counting time and time constant of the exponentially increasing voltage on minimum achievable scanning time is examined by the experiment, and theoretical discussion is made on the basis of the mobility transfer function of particles classified by the DMA, which is related to the mobility shift during the CNC counting time. As a result, accurate measurement of monodisperse particles is completed in about 10 sec. For polydisperse particles, it takes only about ...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.