Abstract
We describe our experience with a quality-assessment scheme for Glucometer users in General Practice based on simultaneous blood samples taken on filter paper strips and posted to the laboratory. The accuracy of results from these blood glucose meters outside the laboratory was below generally accepted laboratory standards; 30% of results fell outside +/- 20% of the laboratory value. Only 58% of the filter paper strips sent out were returned for analysis, suggesting a lack of awareness of the benefits and importance of such a scheme by the G.P. Glucometer users.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.