Abstract
The performance of the newly developed implantable centrifugal blood pump was investigated in vitro. The pump was developed with the end goal of building a versatile system that includes a left ventricular assist system with an internal secondary battery or an implantable biventricular assist system with two implantable blood pumps. The hydrodynamic characteristics and efficiency of the blood pump were evaluated, and the mechanical damage to the blood caused by the blood pump was assessed through a hemolysis test using fresh goat blood. The pump could generate 120 mm Hg at a flow rate of 5 L/min and a motor speed of 2,500 rpm. The electric input power to the pump was approximately 5 watts under these working conditions. The hemolysis caused by the pump was a bit higher than that by the former model, but stayed within an acceptable range. Performance of the pump in vitro was considered sufficient for a left ventricular assist device, although further design improvement is necessary in terms of hemolysis and system efficiency to improve biocompatibility of the pump.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: ASAIO journal (American Society for Artificial Internal Organs : 1992)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.