Abstract

The design and performances of a newly built electrostatic charge state analyzer constructed to act as a spectrometer for keV/u ions are reported. It consists of two 90∘ curved electrodes enclosed by Matsuda electrodes. This setup was recently tested using Ar9+ and Ar12+ ion beams at an energy of 10 keV per charge unit. This spectrometer achieves a good separation of different charge states formed by electron capture processes during collisions between primary ions and the residual gas. Thanks to these first tests, we have identified up to three different background contributions on the detector that need to be reduced or suppressed.