Abstract

We report the performance of a helium circulation system (HCS) for a magnetoencephalography (MEG) that re-liquefies all the evaporating helium gas using two 1.5 W GM cryocoolers operating at 4.2 K. The MEG with the HCS was used to measure human brain responses for over one and a half years without any noise problems. The noise level is below 10 fT/Hz 1/2 for 2–40 Hz, below 30 fT/Hz 1/2 at 1 Hz, and 200 fT/Hz 1/2 at 50 Hz, which is the power supply frequency. As the amount of liquid helium used decreases less than one percent, the maintenance cost of the MEG becomes less than one-tenth of the previous cost.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.