Abstract

lnitial results of the performance of an experimental dual six-port automatic network analyzer operating in the 2-18-GHz range with thermistor-type power detectors are given. The imprecision in measuring reflection coefficients of one-port devices, or the scattering parameters of two-port devices is 4x10/sup -5/, excluding connector repeatability. At 3 GHz, the imprecision in measuring attenuation varies from 0.0003 dB at low values of attenuation to 0.15 dB at 60 dB. The systematic error in measuring attenuation appears to be Iess than the imprecision. The systematic error in measuring reflection coefficient appears to be less than 0.0004. Additional systematic errors caused by changes in the calibration constants over a 20-week period were observed to be less than 0.003 dB in attenuation and less than 0.002 in reflection coefficient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.