Abstract

The thermal neutron flux can be easily measured in real time by using the scintillator with optical fiber (SOF) detector. However the irradiation damage under high-intensity neutron flux causes deterioration of the SOF detector due to radiation damage to the plastic scintillator in which 6LiF is blended. After irradiating the SOF detector for 4 h (thermal neutron fluence is approximately 2.0×10 13 neutrons/cm 2), the sensitivity of the SOF detector decreased by 3.0%. After irradiating the SOF detector for 2 months (thermal neutron fluence approximately 6.4×10 14 neutrons/cm 2), the sensitivity was reduced to 42% of baseline. Supposing that the thermal neutron fluence is 2×10 12 neutrons/cm 2 on the surface of a patient in a BNCT treatment, the sensitivity of the SOF detector is reduced by approximately 0.3%. This report presents investigations on the deterioration of the SOF detector in irradiation experiments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.