Abstract

By introducing wavelet-denoising (WD) in the traditional tunable diode laser absorption spectroscopy (TDLAS), a novel CH4 detection device based on WD-assisted wavelength modulation technique was experimentally demonstrated. Formulation and detection procedure were presented, and experiments were carried out to prove the normal function on the extraction of 2f signal from severely polluted differential signal with WD algorithm. With the fabricated device, sensing characteristics were measured on CH4 concentration. The response time is less than 8s within the whole detection range of 0–50kppm. By virtue of the suppression on noises with WD, the minimum detection limit (MDL) is dropped from 4 to 1ppm, the maximum detection error is minimized from 6.2% to 3.8% over the concentration range of 4–50kppm, and the decreased Allan deviation (per an average observing time of 9min for 24h) from 0.13 to 0.08ppm suggests enhanced stability. The proposed device also exhibits improved performances compared with our previously reported sensors based on simple direct spectroscopy absorption (DSA) technique and wide-band incandescence wire-source.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.