Abstract

The Alinity i system (Abbott), a recently developed automated immunoassay analyzer, has a compact footprint and a throughput of 200 tests per hour. Here, we present the first performance evaluation of thyroid function test (TFT) on the Alinity i system. We performed precision, linearity, comparison, functional sensitivity, carryover, and reference interval verification of 4 hormones (Thyroid Stimulation Hormone; TSH, total T3, free T4, and total T4) using the reagents provided by the manufacturer following the guidelines of the Clinical Laboratory Standards Institute. The performance of Alinity i was compared to that of the Architect i2000 immunoassay analyzer (Abbott, US). The within-laboratory coefficients of variation for all the hormones were 1.6-3.6%. Linearity was observed for all the hormones over the entire tested analytical range (R2 ≥ 0.99). The results for all the evaluated hormones using Alinity i system strongly correlated (r ≥ 0.978) with those from Architect i2000. Functional sensitivity was lower than the lower limit of the analytical measurement range. Sample carryover was less than 1.0%. TFTs on the Alinity i system showed acceptable performance in terms of precision, linearity, comparison, functional sensitivity, and carryover. Therefore, this system could be a useful laboratory tool for performing TFTs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.