Abstract

The performance of thermal cyclers for polymerase chain reactions (PCR) is of great concern in terms of the reliability of PCR-based assays, particularly when rapid cycling conditions are applied to small volume reactions. In this work, the precision of the temperature controls during rapid thermal cycling was measured in 19 commercial thermal cyclers of 8 different models. The temperatures of test solutions in specific locations in each thermal block were simultaneously monitored at 1 s intervals during thermal cycling. A temperature-sensitive multiplex PCR was run in parallel to assess undesirable PCR results caused by poor temperature control. Under the given conditions (20 s of annealing time and 20 microL reaction volume), a majority of the tested instruments showed prominent curving, undershooting, and/or overshooting in their temperature profiles, which substantially influenced the results of the temperature-sensitive multiplex PCR. Variations between wells were also observed in most instruments. It is strongly hoped that these problems will be addressed by manufacturers and that they will make substantial improvements in the precision and efficiency of thermal cyclers. In the meantime, users of thermal cyclers might be able to avoid unexpected poor outcomes of sensitive PCR-based assays by designing their PCR protocols with these findings in mind.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call