Abstract
ObjectivesPediatric hospitals are always challenged by specimen volumes and thus any innovation in this realm is very welcome. With the introduction of Microslide assay pairs, we aimed to evaluate the analytical performance of the Vitros XT MicroSlide assay pairs on the Vitros XT 7600 compared to single MicroSlides. DesignPerformance characteristics included within-run precision, analytical measurable range, method comparison, and interference verification. We compared six XT MicroSlide pairs on the Vitros XT 7600 with twelve corresponding single slide assays on the Vitros 5600 system. ResultsThe XT MicroSlides on Vitros XT 7600 demonstrated excellent precision, equivalent analytical measurable range, and strong method correlation with single slide assays on Vitros 5600 for most of the assays tested. Within-run CVs of the analytes ranged between 0.32% and 2.93% with between-run CV of less than 8.8% and linearity for all analytes was within the manufacturer's specified range. Interference studies showed comparable effects of hemolysis, lipemia, and bilirubin on both instruments. ConclusionsThe XT MicroSlides are comparable to the single MicroSlide assays with improved efficiency, turnaround times and lower sample volumes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.