Abstract
The Concorde MicroPET Rodent R4 system designed by Concorde Microsystems Inc. is the dedicated positron emission tomography (ECT) for studies of rodents. The scanner has 32 detector rings and a 120mm animal aperture. The transversal field of view(FOV) is 100mm, and the axial FOV is 78mm. The spatial resolution at the center of FOV is smaller than 1.8mm. This paper designs a series of experiments to evaluate the performance parameters (spatial resolution, sensitivity, scatter ets.) of the scanner. In the center FOV the transversal spatial resolution is 1.9mm, the axial spatial resolution is 1.88mm. The absolute sensitivity of the system is 39.88 cps/kBq for an energy window of 250KeV-750KeV. For the same energy window the 2D-reconstructed scatter fractions is 50.6%, the 3D-reconstructed scatter fractions is 32.3%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.