Abstract

At the Department of Physics of the University of Pisa, Italy, a new and fully engineered version of the YAP-(S)PET small animal scanner has been recently installed. The new YAP-(S)PET is able to perform both PET and SPECT studies on small animals. The scanner is made up of four heads: each one is composed of a 4/spl times/4 cm/sup 2/ YAlO/sub 3/:Ce (or YAP:Ce) matrix of 20/spl times/20 elements, 2/spl times/2/spl times/25 mm/sup 3/ each, coupled to a Position Sensitive Photomultiplier (PS-PMT) (Hamamatsu R2486). The four modules are positioned on a rotating gantry. The switching to the SPECT modality is easily made by mounting a high resolution parallel hole (0.6 mm O, 0.15 mm septum) lead collimator in front of each crystal. This paper reports the performance of the system in terms of absolute sensitivity and spatial resolution for both PET and SPECT modalities. The scatter fraction and noise-equivalent count rate (NEC) for a mouse phantom for different energy windows have been measured in PET modality. Images of phantoms and animals are also presented.