Abstract

The Tandem Differential Mobility Analyzer (TDMA) technique coupled with aerosol humidification has been widely used for studying aerosol hygroscopicity. In this study, we evaluate the performance of a commercial Humidified TDMA (BMI HTDMA, Model 3002) with respect to DMA sizing, relative humidity (RH) control, and growth factor (GF) measurements. Unique features of this particular HTDMA include a diffusion-based particle humidifier, a DMA design allowing selection of particles up to 2 μm diameter at only 5600 volts, and the ability to study the complete deliquescence and efflorescence cycle. The sizing agreement between DMA 1 and 2 was within 2% over the 35 to 500 nm diameter range. The measured TDMA responses agreed well with theoretical calculations. The RH control and stability were tested at a suburban field site in Hong Kong. The system achieved RH equilibrium in less than 4 min when changing the RH set point. With indoor temperature changes of less than 1°C per hour, the RH control of the system was...