Abstract

This study evaluated the diagnostic and analytical performances of the Access anti-HBV core antigen (anti-HBc Total) assay on the DxI 9000 Access Immunoassay System (Beckman Coulter Inc.). The multicenter study involved both prospective and retrospective sample collection from non-selected blood donors, hospitalized patients, or presumed anti-HBc Total positive individuals. Fresh/previously-frozen samples were tested with the Access and comparator assays to determine concordance; discrepant samples were tested with a second CE-marked assay. Among the 5983 non-selected fresh blood donor samples deemed anti-HBc Total negative, clinical specificity of the Access assay was 99.58% (95%CI: 99.38–99.72%). Clinical specificity was 99.27% (97.37–99.80%) among 273 anti-HBc Total negative hospitalized patient samples. Clinical sensitivity on 450 anti-HBc Total positive samples was 99.78% (98.75–99.96%). Evaluation in seroconversion panels revealed an average 1.4-day earlier detection versus a comparator assay. The Access assay demonstrated excellent clinical and analytical performances comparable to existing CE-marked anti-HBc Total assays. NCT04904835.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call