Abstract

The present study was carried out with the aim of evaluating the performance of six different aerosol samplers in terms of mass concentration, particle size distribution, and mass fraction for the international size-sampling conventions. The international size-sampling criteria were defined as inhalable, thoracic, and respirable mass fractions with 50% cutoff at an aerodynamic equivalent diameter of 100 μm, 10 μm, and 4 μm, respectively. Two Andersen, four total suspended particulate (TSP), two RespiCon, four PM 10, two DustTrak, and two SidePak samplers were selected and tested to quantitatively estimate human exposure in a carefully controlled particulate matter (PM) test chamber. The overall results indicate that (1) Andersen samplers underestimate total suspended PM and overestimate thoracic and respirable PM due to particle bounce and carryover between stages, (2) TSP samplers provide total suspended PM as reference samplers, (3) TSP samplers quantified by a coulter counter multisizer provide no information below an equivalent spherical diameter of 2 μm and therefore underestimate respirable PM, (4) RespiCon samplers are free from particle bounce as inhalable samplers but underestimate total suspended PM, (5) PM 10 samplers overestimate thoracic PM, and (6) DustTrak and SidePak samplers provide relative PM concentrations instead of absolute PM concentrations.