Abstract

The performance evaluation studies of three indigenously manufactured amperometers were carried out for uranium determination in U3O8 standard solution. Three analysts independently carried out ten determinations each of uranium by Biamperometric end point detection method with each of these amperometers. Two-way ANOVA test was performed on the data using MATLAB-7.0 software to test whether the variation in the observed data is statistically significant. No significant bias in the data produced by three analysts with three independent instruments was observed. The results obtained show that the indigenous amperometers are suitable for uranium determination in fuel materials and for nuclear material accounting, with high precision and accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.