Abstract
In this study, cellular phone network based portable ECG device was developed and three experiments were performed to evaluate the accuracy, reliability and operability, applicability during daily life of the developed device. First, ECG signals were measured using the developed device and Biopac device (reference device) during sitting and marking time and compared to verify the accuracy of R-R intervals. Second, the reliable data transmission to remote server was verified on two types of simulated emergency event using patient simulator. Third, during daily life with five types of motion, accuracy of data transmission to remote server was verified on two types of event occurring. By acquiring and comparing subject's biomedical signal and motion signal, the accuracy, reliability and operability, applicability during daily life of the developed device were verified. Therefore, cellular phone network based portable ECG device can monitor patient with inobtrusive manner.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.