Abstract

BackgroundTo validate the point of care testing (POCT) Trop-I analyzer and compare it with a central laboratory-based chemiluminescence immunoassay, in order to evaluate its performance for use in critical care areas. Moreover, for clinical decision-making, it is imperative to know the extent to which patient stratification will differ based on the analytic method being used. In particular, the aim of this study was to evaluate the analytical performance of the point-of-care analyzer and demonstrate the agreement with the central laboratory measurements in patients presenting to the emergency department (ED) with chest pain and suspected acute coronary syndrome (ACS).MethodsThis cross-sectional study was performed at the section of chemical pathology, department of pathology and laboratory medicine, the Aga Khan University (AKU), Karachi, from October to November 2017. Samples from patients and the quality control material of Trop-I were analyzed for imprecision, linearity, and method comparison on Advia Centaur (Siemens Diagnostics, CA, USA), and the AQT90 FLEX analyzer (Radiometer Medical ApS, Brønshøj, Denmark) with photometric detection at the section of chemical pathology, AKU. Statistical analysis was done using Microsoft Excel (Microsoft Corporation, Washington, United States) and EP Evaluator version 10.3.0.556 (Data Innovations, LLC, VT, US). Quantitative variables were represented in terms of mean ± SD. For precision, the computed SD was compared with allowable random error. Furthermore, Cohen’s kappa was applied to observe the agreement between the two methods.ResultsThe Trop-I Precision study on the POCT analyzer showed a coefficient of variation (CV) of 2.4% using a pooled patient sample with a mean Trop-I of 2.15 ± 0.05 ng/ml. Three standards ranging from 0.034 to 1.316 ng/ml were run in triplicate to verify accuracy and linearity. The allowable systematic error (SEa) was 10.0%. The maximum deviation for a mean recovery from 100% was 4.1%. All three of the mean recoveries were accurate and within the allowable error limits. The results were linear with slope 1.04, intercept 0.0. On a method comparison, Trop-I showed good agreement, yielding a kappa value of 0.95.ConclusionThis study has validated the performance of a POCT Trop-I assay against a central laboratory immunoassay and found acceptable results. POCT assays for cTnI should be implanted in emergency settings to ensure the fast triage of patients with chest pain, as well as timely diagnosis.

Highlights

  • The rapid diagnosis of acute coronary syndrome (ACS) relies pivotally on cardiac troponin assays [1]

  • This study has validated the performance of a point of care testing (POCT) Trop-I assay against a central laboratory immunoassay and found acceptable results

  • POCT assays for cTnI should be implanted in emergency settings to ensure the fast triage of patients with chest pain, as well as timely diagnosis

Read more

Summary

Introduction

The rapid diagnosis of acute coronary syndrome (ACS) relies pivotally on cardiac troponin (cTn) assays [1]. An evaluation of worldwide practices shows that most of the central hospital laboratories find it difficult to comply with these guidelines of the rapid availability of cTn results for optimum patient management [4]. How to cite this article Ahmed S, Jafri L, Raheem A, et al (March 11, 2019) Performance Evaluation of Cardiac Troponin I Assay: A Comparison Between the Point-ofcare Testing Radiometer AQT90 FLEX and the Central Laboratory Siemens Advia Centaur Analyzer. To validate the point of care testing (POCT) Trop-I analyzer and compare it with a central laboratory-based chemiluminescence immunoassay, in order to evaluate its performance for use in critical care areas. The aim of this study was to evaluate the analytical performance of the point-of-care analyzer and demonstrate the agreement with the central laboratory measurements in patients presenting to the emergency department (ED) with chest pain and suspected acute coronary syndrome (ACS)

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.