Abstract

BackgroundThis study aimed to evaluate the performance of a preclinical PET insert in three configurations: as a stand-alone unit outside the MRI bore, inside the bore of a cryogen-free 3T MRI and, finally, while performing simultaneous PET/MRI studies.MethodsThe PET insert consists of two rings of six detectors, each detector comprising 8 × 12 SiPMs reading out dual offset layers of pixelated LYSO crystals with a 1.4-mm pitch. The inner diameter is 60 mm, transaxial field of view (FoV) 40 mm and axial FoV 98 mm. Evaluation was based on NEMA NU 4-2008 guidelines with appropriate modifications. Spatial resolution and sensitivity were measured inside and outside the MR bore. Image quality, count rate and quantitative performance were measured in all three configurations. The effect of temperature stability on PET sensitivity during fast spin echo sequences was also evaluated. B0 field homogeneity and T1 and T2 relaxation times were measured using a water-filled phantom, with and without simultaneous PET operation. Finally, PET and MRI scans of a mouse injected with 10 MBq [18F]NaF and a mouse injected with 16 MBq [18F]FDG were performed in sequential and simultaneous modes.ResultsPeak absolute sensitivity was 10.15% with an energy window of 250–750 keV. Absolute sensitivity values outside and inside the MR bore with MR idle agreed to within 0.1%. Outside the MR bore, spatial resolution was 1.21/1.59 mm FWHM (radial/tangential) 5 mm from the centre of the FoV which compared well with 1.19/1.26 mm FWHM inside the MR bore. There were no substantial differences between all three scan configurations in terms of peak NEC rate (175 kcps at 17 MBq), scatter or random fractions. Uniformity and recovery coefficients were also consistent between scanning modes. B0 field homogeneity and T1 and T2 relaxation times were unaltered by the presence of the PET insert. No significant differences were observed between sequential and simultaneous scans of the animals.ConclusionsWe conclude that the performance of the PET insert and MRI system is not significantly affected by the scanning mode.

Highlights

  • The combination of positron emission tomography (PET) and magnetic resonance imaging (MRI) has several advantages over other hybrid imaging approaches for preclinical [1] and clinical research

  • No significant differences were observed between spatial resolution measured with the PET insert positioned inside the MR bore and outside the bore

  • The small-animal PET insert evaluated in this study was designed to allow simultaneous imaging with preclinical 3T and 7T MRI systems, while retaining the flexibility to be used as a stand-alone unit on the benchtop

Read more

Summary

Introduction

The combination of positron emission tomography (PET) and magnetic resonance imaging (MRI) has several advantages over other hybrid imaging approaches for preclinical [1] and clinical research. There are multiple ways in which the complementary information from PET and MRI can be combined for enhanced quantitative accuracy and reliability, including joint estimation of image and/or kinetic parameters [4]. This versatility means there are many parameters to optimise, including when to acquire the data simultaneously versus sequentially in order to solve important biological and clinical problems. To enable simultaneous PET/MRI, numerous challenges had to be overcome [5], including the effects of magnetic fields and radiofrequency (RF) signals on PET detectors [6,7,8]. This study aimed to evaluate the performance of a preclinical PET insert in three configurations: as a stand-alone unit outside the MRI bore, inside the bore of a cryogen-free 3T MRI and, while performing simultaneous PET/MRI studies

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call