Abstract

To evaluate the analytical characteristics of anovel high-sensitivity cardiac troponin T (hs-cTnT) test onthe automatic light-initiated chemiluminescent assay (LiCA®) system, and validated its diagnostic performance for non-ST-segment elevation myocardial infarction (NSTEMI). Studies included an extensive analytical evaluation and established the 99th percentile upper reference limit (URL) from apparently healthy individuals, followed by a diagnostic performance validation for NSTEMI. Sex-specific 99th percentile URLs were 16.0 ng/L (1.7 % CV: coefficient of variation) for men (21-92 years) and 13.4 ng/L (2.0 % CV) for women (23-87 years) in serum, and 30.6 ng/L (0.9 % CV) for men (18-87 years) and 20.2 ng/L (1.4 %CV) for women (18-88 years) in heparin plasma. Detection rates in healthy individuals ranged from 98.9 to 100 %. An excellent agreement was identified between LiCA® and Elecsys® assays with a correlation coefficient of 0.993 and mean bias of-0.7 % (-1.8-0.4 %) across the full measuring range, while the correlation coefficient and overall bias were 0.967 and-1.1 % (-2.5-0.3 %) for the lower levels of cTnT (10-100 ng/L), respectively. At the specific medical decision levels (14.0 and 52.0 ng/L), assay difference was estimated to be <5.0 %. No significant difference was found between these two assays in terms of area under curve (AUC), sensitivity and specificity, negative predictive value (NPV) and positive predictive value (PPV) for the diagnosis of NSTEMI. LiCA® hs-cTnT is a reliable 3rd-generation (level4) high-sensitivity assay for detecting cardiac troponin T. The assay is acceptable for practical use in the diagnosis of NSTEMI.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call