Abstract

The DoPET scanner is a compact positron emission tomography (PET) device. It has been developed for monitoring the range of charged particles during therapy with hadron beams. Previous works have focused on the development and upgrade of the device and on data analysis. In this paper, a full performance characterization of the DoPET system in terms of the energy resolution, spatial resolution, sensitivity, uniformity, and noise equivalent count rate is reported. All measurements refer to an adapted version of the National Electrical Manufacturers Association (NEMA) NU 4 - 2008 protocol, which was written originally for small animal PET systems. Since DoPET is a dual head planar system, it requires a modified characterisation procedure with respect to those described for ring geometries as in the NEMA NU 4 - 2008 protocol. The presented procedure may be of interest for any other PET system with a similar geometry as DoPET.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.