Abstract
The equipment is not sterile when the health equipment is used repeatedly. Thus, it is at risk of infection. Data published by WHO reveals that one out of every ten hospital patients gets an infection. In developed countries, the number of infected hospital patients reaches 30%, while in developing countries it is at least 2-3 higher. One way to reduce the occurrence of infection is to sterilize medical equipment through a dry heat sterilization process at 1700 C for 60 minutes. The high price of sterilizers means that many health centers and clinics do not have this equipment. The aim of this study was to design a low-cost dry heat type health equipment sterilizer using fuzzy logic as a controller and evaluate the performance of the controller. Regarding to the research result, it indicated that the controller performance evaluation was very good with the highest overshoot value of 1.1%, the steady state error value was 1.1%, and the fastest rise time was 240 seconds.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.