Abstract
Detecting presence of male DNA in a mixture is still a challenge in forensic DNA typing. The study presents performance comparison of the two most recent Y-STR systems available for forensic case work, Y filer plus (YFP) and PowerPlex Y23 (PP Y23) multiplex system. Comparative study was conducted by using full and half reaction volumes of both the systems. On the basis of the work done on more than 200 case samples of sexual assault, PP Y23 multiplex system was found to be more promising than YFP. In the experiments conducted with lesser reaction volumes, it was observed that even 8 μL PCR reaction mix of PP Y23 was able to detect 16 pg of male DNA; however, with YFP it was possible with 100 pg. More balanced Y-STR profiles were observed with low male DNA with PPY23.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Forensic Science International: Genetics Supplement Series
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.