Abstract

The performance characteristics of four compact, room-temperature detectors - two scintillators and two semiconductor detectors - were studied. All are commercially-available detectors. The two scintillators were a O13 mm /spl times/ 13 mm lanthanum chloride [LaCl/sub 3/(Ce)] detector and a O25 mm /spl times/ 25 mm sodium iodide [NaI(Tl)] detector. The two semiconductor detectors were a 10 mm /spl times/ 10 mm /spl times/ 3 mm cadmium zinc telluride (CZT) detector with a coplanar gridded anode and a 5 mm /spl times/ 5 mm /spl times/ 5 mm CZT detector with an extended cathode. The efficiency, resolution, and peak shape of these devices are compared. Since LaCl/sub 3/(Ce) is a relatively new commercial scintillator material, additional information on the performance of this detector is presented. Specifically, the impact of naturally-occurring radioactive /sup 138/La and additional contamination from alpha-emitting radionuclides on the background measured with this detector are discussed.