Abstract

BackgroundThe aim of this study is to conduct physical performance evaluation on the uMI550 whole-body PET/CT system according to the NEMA NU2-2018 standard.MethodsAccording to the NEMA NU2-2018, spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses and random corrections, image quality, and timing resolution were evaluated. Spatial resolution was measured by using a 22Na point source. System sensitivity was measured by inserting an 18F line-source in six concentric aluminum sleeves with varying diameters. Scatter fraction, count-rate performance, accuracy of count loss, and timing resolution were all calculated by analyzing dynamically acquired data of an 18F line-source inside a polyethylene cylinder in 20 cm diameter and 70 cm length. Image quality was assessed using a NEMA IEC body phantom with a 4:1 ratio of activity concentration of spheres to the warm background. Additionally, three patient studies were performed, with one brain scan and two whole-body scans, separately. The patient images were evaluated to get a visual first impression of uMI550.ResultsThe tangential, radial, and axial spatial resolutions were measured as 2.91 mm, 2.98 mm, and 2.97 mm FWHM, respectively, at 1 cm radial offset. The total system sensitivity to line source at center was 10.24 cps/kBq. A NECR peak was measured as 124.4 kcps at 18.85 kBq/mL. The scatter fraction at NECR peak was 36.65%, and the maximum count-rate error at and below NEC peak was 1.55%. Contrast recovery coefficients were from 46.5 (10 mm) to 83.9% (37 mm). The timing resolution was measured as 372 ps at low count rate.ConclusionNEMA NU-2 2018 testing was performed on the new SiPM-based uMI550 PET/CT system. The uMI550 shows a high-spatial resolution of less than 3 mm and a good timing resolution of 372 ps. It shows clinical significances on improving potentially diagnostic ability on small lesions.

Highlights

  • The aim of this study is to conduct physical performance evaluation on the uMI550 whole-body Positron emission tomography/computed tomography (PET/CT) system according to the National Electrical Manufacturers Association (NEMA) NU2-2018 standard

  • NEMA measurements Spatial resolution At 1 cm radial offset, the tangential, radial, and axial spatial resolutions were measured as 2.91 mm, 2.98 mm, and 2.97 mm FWHMs, respectively; at 20 cm radial offset, the Radial distance resolutions were degraded to 4.08 mm, 4.14 mm, and 3.13 mm, respectively

  • Scatter fraction and count-rate performance As shown in Fig.2 a, the count rates of noise equivalent count (NEC), prompts, trues, randoms, and scatter varied with radioactivity concentration are plotted as curves, and the peak NEC rate is measured as 124.4 kcps at 18.85 kBq/mL activity concentration As shown in Fig. 2b, the scatter fraction at peak NEC rate is 36.65%

Read more

Summary

Introduction

The aim of this study is to conduct physical performance evaluation on the uMI550 whole-body PET/CT system according to the NEMA NU2-2018 standard. The lutetium oxyorthosilicate (LSO) crystals [4] advanced time-of-flight (TOF) technology to stimulate the development of commercial TOF-PET scanner [5,6,7,8]. The uMI550 PET/CT (Shanghai United Imaging Healthcare Co., Ltd.) (hereinafter referred to as uMI550) was developed by using silicon photomultiplier (SiPM)-based detectors with 16.3 mm LYSO crystals. NEMA NU 2-2018 (published by the National Electrical Manufacturers Association (NEMA)) [15] allows a reproducible measurement standard for evaluating physical performance of PET systems. The purpose of this study was to evaluate uMI550 physics performance, including spatial resolution, sensitivity, scatter fraction, noise equivalent count rate (NECR), image quality, accuracy of count losses and random correction, and timing resolution.

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call