Abstract
Background: Serum cortisol measurements are used to help diagnose conditions of cortisol deficiency and excess. Automated, nonisotopic immunoassays are used in most clinical laboratories. Methods: Linearity, imprecision, and method comparison studies for serum cortisol assays were performed with the Access®, Advia Centaur®, AxSYM®, Elecsys® 2010, and IMMULITE® 2000 analyzers. Results: All methods demonstrated acceptable linearity with the AxSYM method showing the least deviation from target values. The Elecsys 2010 method was the most precise for all three levels of quality control material. Method comparison studies demonstrated good concordance among methods except for two patient samples that gave results on the Elecsys 2010 that were significantly higher than the other methods. Differences in assay calibration were observed. Conclusions: Overall, all methods performed well. Additional calibration standardization efforts are required to truly harmonize the results from each assay. The Elecsys 2010 method is less specific than other automated methods based on two discordant patient results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.