Abstract

Serum concentrations of cancer antigen (CA) 19-9 can be useful in monitoring response to therapy in pancreatic cancer. The objective of this study was to evaluate 5 automated CA 19-9 assays: ARCHITECT <italic>i</italic>2000 (Abbott Diagnostics, Abbott Park, IL), ADVIA Centaur (Bayer Diagnostics, Tarrytown, NY), UniCel DxI 800 (Beckman Coulter, Fullerton, CA), IMMULITE 2000 (Diagnostic Products, Los Angeles, CA), and Elecsys E170 (Roche Diagnostics, Indianapolis, IN). All methods were evaluated for limit of detection, linearity, imprecision, method comparison, and reference intervals. Limit of detection results were all below 2.0 kU/L and met the manufacturers’claims. Linearity had deviation from target values that ranged from 4.5% to 26.7%. All methods showed acceptable imprecision with total coefficients of variation less than 8%. Method comparison by Passing-Bablok analysis resulted in slopes ranging from 1.00 to 2.06 and correlation coefficients of 0.85 to 0.98. Between 97.6% and 99.2% of results from healthy volunteers were less than 35 kU/L. All methods show acceptable analytic

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.